Heated Sample Lines Rental with Temperature Controller

Heated Sample Line RentalCleanAir Instrument Rental heated sample lines feature an electrically heat traced and insulated 3/8″ PTFE (.030 Wall) sample line with a stainless steel over braid and stainless steel tube ends, a 1/4 PTFE (.040 wall) calibration line, 3 pin Amphenol power connector, and 2 type K thermocouple plugs. Protected by a durable scuff resistant extruded polyurethane jacket. *Temperature controller sold separately. *Heated Sample Lines are not self limiting; a temperature controller is required. Custom lengths, voltage, and configuration available *Heavy Duty Heated Sample Lines are manufactured to have a continuous operating temperature of 400°F (~204°C) at -20°F ambient temperature.

Rentals orders come with a Temperature Controller.  New or Used Heated Sample Lines require an additional purchase.  Temperature controllers are available here.

Part Number
Voltage
Watts
Length FT
Amp.
0723-10R
120
300
10′ Heated Sample Line
2.73
0723-25R
120
750
25′ Heated Sample Line
6.82
0723-50R
120
1500
50′ Heated Sample Line
13.64
0723-100R
120
2500
100′ Heated Sample Line
22.73

Heated Sample Lines are used with:

Heavy Duty versus Standard Heated Sample Line

      • Note: Heavy Duty Heated Sample Lines HD are manufactured to have a continuous operating temperature of 400°F (~204°C) at an ambient temperature of -20°F.
      • Note: Standard Heated Sample Lines are manufactured to have a continuous operating temperature of 400°F (~204°C) at an ambient temperature of 0°F.

Heated Sample Line Features:

  • Rated for 400 º F continuous operation at -20 º F ambient temperature
  • Triple insulation maintains temperature of line with less power consumption. (Heavy Duty)
  • Lower resistance heaters requires less power to heat lines.
  • A backup type K thermocouple to prevent project delays in the case of primary thermocouple failure in the field.
  • Stainless steel over braid and stainless steel tube ends for the sample line to prevent abrasive failure in the field. Optional stainless steel over braid for protection of calibration line from abrasive failure in the field.
  • Durable extruded poly-urethane jacket for protection of sample/calibration lines from abrasive failure in the field.
  • Inert PTFE sample and calibration lines provide more accurate results by eliminating potential bias from other materials. PTFE is also more corrosion resistant than tygon, and can be washed with acetone without degradation.

No additional documents available for this product

CleanAir Engineering provides world-class service, support, and technical expertise alongside every instrument rental. All air monitoring equipment is inspected or calibrated before shipping from one of our four technical centers. Each rental includes additional accessories and components, providing added convenience and reliability to ensure the success of your project. Contact us for more information.