Mobile FTIR (CEM) System

 CleanAir® designed this mobile FTIR system to sample unattended for extended durations.  Data can be accessed remotely via a cellular modem or ethernet conection.  An autofill system for the liquid nitrogen cooled detector can be included, which provides up to 1 week of LN2.  Automatic calibration checks are controlled with an ESC datalogger.  The temperatures of all of the sample delivery system are monitored as well.  If a power outage or problem occurs then the sample will bypass the FTIR.
 

Overview

Power: 
480 VAC, 20 Amps
Shipping Weight: 
>350 lbs

Helpful Hints

Helpful Hints: 
 Training on the MKS 2030 FTIR and on this system is required with a rental of this system.   Liquid nitrogen must be supplied by a local vendor. The system requires a minimum of 80 psi.  The cabinet requires a 10KVA single phase transformer, 480V primary, 240/120V secondary.  The cabinet needs 2 legs of 480V 3phase and a ground connection to establish neutral for the secondary. Our cabinets have a fused disconnect switch on the outside (20A fuses) with a 100 foot 10/3 SO cord attached.  We also attach a 4 wire 60A plug that mates with common welder outlets but can be removed to wire directly.

Download Detailed Specifications

Detailed Specifications: 

Photos

Featured Photos: