Measure trace levels of carbon monoxide (CO) in ambient air with the Thermo Scientific Model 48i-TLE CO Analyzer (Enhanced Trace Level), which utilizes gas filter correlation and NDIR technology. The enhanced sample conditioned temperature stabilization and auto-zeroing of the Model 48-TLE result in a low detection limit for the measurement of carbon monoxide.

SKU:
Thermo 48 iTL

Measure trace levels of carbon monoxide (CO) in ambient air with the Thermo Scientific Model 48i-TLE Enhanced Trace Level CO Analyzer, which utilizes gas filter correlation and NDIR technology. The enhanced sample conditioned temperature stabilization and auto-zeroing of the Model 48-TLE result in a low detection limit for the measurement of carbon monoxide.

Using gas filter correlation technology, the Thermo 48i TLE measures the amount of carbon monoxide in the air, based on the principle that carbon monoxide (CO) absorbs infrared radiation at a wavelength of 4.6 microns. Because infrared absorption is a nonlinear measurement technique, the 48i transforms the captured signal into a linear output. The instrument uses an exact calibration curve to accurately linearize the instrument output over any range up to a concentration of 1,000 ppm.

Features

  • Gas filter correlation technology
  • Standard dual range and auto ranges
  • Self aligning optics
  • Temperature and pressure correction
  • User settable alarm levels for concentration

Thermo i Series Features

  • Ethernet connectivity for efficient remote access
  • Enhanced user interface with one button programming and large display screen
  • Flash memory for increased data storage and user downloadable software
  • Enhanced electronics design optimizes product commonality
  • Improved layout for easier accessibility to components

Specifications

Detectable Analytes Carbon Monoxide
Flow Rate 0.5 L/min.
Inputs 16 Digital Inputs (Standard), 8 0 to 10 VDC Analog Inputs (Optional)
Linearity ±1% full scale
Response Time 60 sec. (30 sec. average time)
Span Drift ±1% full scale
Temperature (English) Operating 68°F to 86°F
Temperature (Metric) Operating 20°C to 30°C
Certifications/Compliance ETL, CSA, UL, CE; U.S. EPA (RFCA-0981-054), UKCA
Depth (English) 23 in.
Depth (Metric) 584 mm
Description Model 48i-TLE Enhanced Trace Level CO Analyzer
Detection Limit 0.04 ppm (30 second averaging time)
Height (English) 8.62 in.
Height (Metric) 219 mm
Measurement Range Preset: 0 to 50, 100, 200, 500, 1000, 2000, 5000, 10000 and 20000 ppm; 0 to 50, 100, 200, 500, 1000, 2000, 5000, 10000 and 20000 mg/m3; Custom: 0 to 50 to 20000 ppm, 0-50 to 20000 mg/m3
Outputs Selectable Voltage, RS232/RS485, TCP/IP, 10 Status Relays, and Power Fail Indication (standard); 0 to 20 or 4 to 20mA Isolated Current Outout (optional)
Voltage 100/115 VAC, 220/240 VAC
Weight (English) 55 lb.
Weight (Metric) 25 kg
Width (English) 16.75 in.
Width (Metric) 425 mm
Zero Drift <0.1ppm

No additional documents available for this product

CleanAir Engineering provides world-class service, support, and technical expertise alongside every instrument rental. All air monitoring equipment is inspected or calibrated before shipping from one of our four technical centers. Each rental includes additional accessories and components, providing added convenience and reliability to ensure the success of your project. Contact us for more information.

CleanAir Engineering offers full-service calibration or repair for this item. To learn more, contact a technician today at: 1-800-553-5511. Or, complete our repair request form to begin your service request today! Most out-of-calibration or out-of-service instruments can be shipped to any one of our four technical service centers, in Chicago, Houston, Pittsburgh, or Marseille, France.